Model Number 214646
Device Problems Dull, Blunt (2407); Mechanical Jam (2983)
Patient Problem Not Applicable (3189)
Event Date 02/25/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Device was used for treatment, not diagnosis.Udi: (b)(4).The lot number is unknown.
 
Event Description
It was reported by affiliate via email that during procedure, the cordcutter ea could not cut the sutures, the scissor got ¿stuck¿.No patient consequences, there was a surgical delay of 10 minutes reported.The device is available to be returned for evaluation.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: according to the information provided, it was reported that during a procedure, the cordcutter *ea could not cut the sutures, the scissor got ¿stuck¿.The complaint device is not being returned, therefore unavailable for a physical evaluation.Since the complaint device was not returned, we cannot determine a root cause for the reported failure.If the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.Given that no lot number was provided, a manufacturing record evaluation (mre) review cannot be performed.If the lot number becomes available, the mre review will be performed. at this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
SUTURE CUTTER
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9901205
MDR Text Key205324237
Report Number1221934-2020-01018
Device Sequence Number1
Product Code FZT
UDI-Device Identifier10886705004270
UDI-Public10886705004270
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 03/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/30/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number214646
Device Catalogue Number214646
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/29/2020
Patient Sequence Number1
-
-