The reason for this revision surgery was reported as an infection.The previous surgery and the surgery detailed in this event occurred 1.3 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the implant device history records (dhr) shows that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were non-conforming material report (ncmr) associated with the main product that may have contributed to the reported event.The devices were within its expiration date at the time of the previous surgery.Non-conforming report history: item: 114822, ncr 12201233, work center 91050, op.Seq.0165, clean and prep, 1 part scrap.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this evaluation that indicate the reported devices were the source or had a direct connection with the patient's infection.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the infection or inhibited the patient's immune system.There are multiple factors that may contribute to an infection that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
|