LIVANOVA DEUTSCHLAND GMBH S5 ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-80-00
Device Problem Pumping Problem (3016)
Patient Problem No Patient Involvement (2645)
Event Date 03/03/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 roller pump.The incident occurred in (b)(6).A review of the dhr could not identify any deviations or nonconformities relevant to the issue, a livanova field service representative was dispatched to the facility to investigate the device and was able to confirm the reported issue.He replaced the motor end stage board and motor control board to solve the malfunction.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The most likely root cause of the issue is a faulty connections between the two boards due to a defective involved components.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova received a report that a s5 roller pump displayed a motor control failure error message during the setup.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9910097
MDR Text Key197335894
Report Number9611109-2020-00225
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
V
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 03/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-80-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/03/2020
Initial Date FDA Received04/01/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/11/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-