• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP SYSTEM STEERABLE GUIDE CATHETER; VALVE REPAIR Back to Search Results
Model Number SGC0301
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Embolism (1829); Thrombosis (2100); Atrial Perforation (2511)
Event Date 03/10/2020
Event Type  Injury  
Manufacturer Narrative
The device will not be returned for evaluation, the device was reportedly discarded.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.Na.
 
Event Description
This is filed to report atrial perforation, thrombus, and medical intervention.It was reported that this was a mitraclip procedure to treat mixed mitral regurgitation (mr) with a grade of 4.It was noted thrombus found on a previously implanted lead.After successful clip-deployment, the steerable guide catheter (sgc) was removed; however, an atrial septal defect (asd) was closed with a closure device due to the thrombus possibly migrated to the left side of the heart and embolized.The thrombus was not resolved.Two clips were implanted, reducing mr to 2.There was no reported clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot.In this case, there was no reported device malfunction associated with the steerable guide catheter.Based on the information provided a conclusive cause for the reported thrombosis cannot be determined.The reported atrial perforation is the result of procedural conditions.The reported embolism is the result of the thrombosis.The reported patient effects of thrombosis, atrial perforation, and embolism, as listed in the mitraclip system instructions for use, are known possible complications associated with mitraclip procedures.There is no indication of a product issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP SYSTEM STEERABLE GUIDE CATHETER
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9911415
MDR Text Key186745837
Report Number2024168-2020-03205
Device Sequence Number1
Product Code DRA
UDI-Device Identifier08717648216824
UDI-Public08717648216824
Combination Product (y/n)N
PMA/PMN Number
K161985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 04/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/08/2020
Device Model NumberSGC0301
Device Catalogue NumberSGC0301
Device Lot Number90809U122
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/10/2020
Initial Date FDA Received04/01/2020
Supplement Dates Manufacturer Received04/06/2020
Supplement Dates FDA Received04/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
2 IMPLANTED MITRACLIPS
Patient Outcome(s) Required Intervention;
Patient Weight120
-
-