Model Number 130790030
Device Problem Difficult to Advance (2920)
Patient Problem No Code Available (3191)
Event Date 03/16/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
(b)(6) drilled and measured hole for metaglene screw as per surgical technique.He was unable to screw the screw in so it was protruding outwards.He red rolled and remeasured but still unable to screw in the screw so it was flush to the metaglene.In the end he opened another screw size 30 and was able to screw it straight in.Original screw was faulty.No adverse injury to patient.Procedure delayed by a few minutes to redrill and retry to instill.The screw in question has been sent for sterilization and can then be collected for investigation from theaters at the hospital.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy as not synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide complaint database search found no other reported incident(s) against the provided product/lot combination(s) since release for distribution.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND SCREW LOCK D4.5X30MM
Type of Device
DELTA XTEND IMPLANTS : SCREWS
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key9911821
MDR Text Key190339907
Report Number1818910-2020-09925
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027843
UDI-Public10603295027843
Combination Product (y/n)N
PMA/PMN Number
K062250
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 03/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number130790030
Device Catalogue Number130790030
Device Lot Number5328058
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/16/2020
Initial Date FDA Received04/01/2020
Supplement Dates Manufacturer Received05/26/2020
Supplement Dates FDA Received05/27/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-