On (b)(6) 2020, a 5/4 amplatzer duct occluder was selected for implant, but was determined to be too small to treat the defect.The device was not released from the cable and was safely removed from the patient.A 6/4 amplatzer duct occluder (lot: 7002193) was then selected for implant, but upon deployment the device deformed into a cobra shape.The device was not released from the cable and was removed from the patient.A second 6/4 amplatzer duct occluder was then selected and was successfully implanted.There are no known complications with this case.There was no clinically significant delay in procedure and the patient remained hemodynamically stable throughout the procedure.
|
Additional information for: g4, g7, h2, h3, h6, and h10.The reported event of device deformation could not be confirmed.The investigation confirmed the device met visual and functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined.
|