• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES¿ ULTRA IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES¿ ULTRA IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1600-04
Device Problems Output below Specifications (3004); Appropriate Term/Code Not Available (3191)
Patient Problem No Information (3190)
Event Date 03/04/2020
Event Type  malfunction  
Event Description
The recipient reportedly experienced decreased performance.The recipient's device was explanted.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed silicone damage on the top and bottom cover prior to receipt.These are believed to have occurred during revision surgery.The photographic imaging inspection revealed an electrode was broken within the electrode pocket.The device passed the electrical tests performed.Advanced electrical testing performed on the device showed lower than typical range on some of the electrodes.It is the opinion of advanced bionics explant review board that some of these electrode test results should be considered as failures.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Programming adjustments were made, however, the issue did not resolve.The recipient was re-implanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key9914048
MDR Text Key187204960
Report Number3006556115-2020-00248
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016841002
UDI-Public(01)07630016841002(11)160912(17)190831
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 02/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date08/31/2019
Device Model NumberCI-1600-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age30 YR
-
-