Updated sections: d10, g4, g7, h2, h3, h6, h10 trackwise # 301532 the device was returned to the factory on (b)(6)2020.An investigation was conducted on (b)(6)2020.A visual inspection was conducted.The c-ring & cannula handle appeared intact.Signs of clinical use and evidence of blood was observed.Blood was observed on the tubes as well as the btt.The end of the insufflation tube was observed to be damaged.It appears as if tape residue was observed on the end of the insufflation tube.The distal end of the insufflation tube was observed to be taped at the end.We are unable to determine when the tape was put on the tube.However, based on the returned condition of the device, we are unable to test the device for the reported failure due to the tape.However, the reported failure ¿no flow¿ was confirmed as well as the analyzed failure, "break- tube damaged".The dhr.Shop floor paperwork was reviewed.The vendor certifies that this device lot conforms to all applicable product specifications.
|