MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problems Overheating of Device (1437); Smoking (1585)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/06/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A review of dhr could not identify any deviations or nonconformities relevant to the issue.The plugs of all heater-cooler devices in use at the hospital had been replaced last year following to the first occurrence of the event (reported under mfr report number 9611109 -2018 -01412).One of the plugs had returned to the manufacturer site for investigation and results revealed that the diameter of the pins of the plug was within the standards.Therefore, it is very likely that the root cause for the overheating was not the plug but the wall socket.
 
Event Description
Livanova received a report stating of a heater-cooler system 3t power supply plug overheating.Smoke was noticed to be coming out from the wall socket.The event was detected during a procedure.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9921375
MDR Text Key186325263
Report Number9611109-2020-00238
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeAU
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 04/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/06/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/04/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-