• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD CONNECTA 3-WAY STOPCOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V. BD CONNECTA 3-WAY STOPCOCK Back to Search Results
Catalog Number 394602
Device Problem Leak/Splash (1354)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/10/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: a device history record review was completed by our quality engineer team for provided lot number 9184190.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.There are quality controls currently in place to detect this type of defect during the production process.Further action has not been determined necessary at this time.Our quality team will continue to monitor the manufacturing process for this defect and other emerging trends.Investigation conclusion: bd was not able to confirm the customer¿s indicated failure mode.Bd was not able to duplicate the failure mode in the manufacturing process.The maintenance records were reviewed and during the manufacture of this lot no adjustments were performed that contributes to this failure mode.Quality records have been consulted for tracking and trending purposes but issues like this are not detected which means pretty low occurrence.We will keep monitoring the manufacturing process and in case any emerging trend is detected, further actions will be taken if necessary.Process fmea rm5785 and eura eurap2053001 were reviewed and there are proper controls in place to detect product malfunctions.Root cause description: based on investigation results to date, root cause for manufacturing process cannot be determined.Rationale: based on an evaluation of severity and frequency it was determined that no corrective action is required at this time also the root cause was not identified, therefore, corrective and preventive actions were not implemented.
 
Event Description
It was reported that the bd connecta¿ 3-way stopcock leaked during use.The following information was provided by the initial reporter, translated from chinese to english: "leakage occured when the connecta was used.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD CONNECTA 3-WAY STOPCOCK
Type of Device
STOPCOCK
Manufacturer (Section D)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX 
Manufacturer (Section G)
BD INFUSION THERAPY SYSTEMS INC. S.A. DE C.V.
periferico luis donaldo
colosio no. 579
nogales
MX  
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652341
MDR Report Key9924286
MDR Text Key195887353
Report Number9610847-2020-00123
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date06/30/2022
Device Catalogue Number394602
Device Lot Number9184190
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/23/2020
Initial Date FDA Received04/03/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/15/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-