• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ZERO TIP; DISLODGER, STONE, BASKET, URETERAL, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ZERO TIP; DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number M0063901050
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/11/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a zero tip stone retrieval baket was to be used in a procedure on (b)(6) 2020.According to the complainant, during preparation, a black stain was noted on the "pb".No patient complications were reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
 
Manufacturer Narrative
H10 block h6: device code 2944 captures the reportable event of "black stain was noted on the pb".Block h10: the zero tip basket was returned in its original unopened pouch.Visual analysis found the heat seal ran throughout the entire length of the pouch without any breakage.Black stains was observed in the front section of the pouch outside the sterile area.No other issues were identified.Based on all available information, it is most likely that handling and manipulation of the device during transport/storage contributed to the observed issue.It is also probable that how the device was handled and manipulated during the processes of shipping/receiving, storage, or packing inside of the generic bag could have contributed to the observed issue.Therefore, the most probable root cause is cause traced to transport/storage.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.
 
Event Description
It was reported to boston scientific corporation that a zero tip stone retrieval baket was to be used in a procedure on (b)(6) 2020.According to the complainant, during preparation, a black stain was noted on the "pb".No patient complications were reported as a result of this event.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZERO TIP
Type of Device
DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key9929721
MDR Text Key191422751
Report Number3005099803-2020-01513
Device Sequence Number1
Product Code FFL
UDI-Device Identifier08714729469643
UDI-Public08714729469643
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/09/2022
Device Model NumberM0063901050
Device Catalogue Number390-105
Device Lot Number0025021773
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/21/2020
Initial Date Manufacturer Received 03/11/2020
Initial Date FDA Received04/06/2020
Supplement Dates Manufacturer Received05/21/2020
Supplement Dates FDA Received06/16/2020
Patient Sequence Number1
-
-