Device evaluation by manufacturer: a distributor engineer visited the customer to address the reported event.The distributor engineer found the home sensor of the sorter was dirty.Fse resolved the complaint by cleaning the sensor.The aia-900 analyzer is operational.There was no further action required by distributor engineer.A 13-month complaint history review and service history review and keyword search for similar complaints was performed for aia-900, serial number (b)(4).There were two (2) similar complaints identified during the searched period, which includes this event.The aia-900 operator's manual under section 11 - maintenance provides detailed information on analyzer maintenance.The most probable cause of the reported event was due to dirty sorter home sensor.
|
A customer reported getting the home sensor of the sorter was not found failure on the aia-900 analyzer.A distributor engineer was dispatched to address the reported event, which resulted in a delay of estradiol (e2), progesterone ii (prog ii), follicle stimulating hormone (fsh) and cardiac troponin i (ctnl2) patient results.There was no indication of patient intervention or adverse health consequences due to the delay in reporting.
|