Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.A field service engineer (fse) was at the customer's site to further investigate.Fse could not duplicate reported issue since all qc results were in acceptable ranges.Fse noticed a few intermittent dl flags on results and decided to clean the detector and replace the 3-way substrate valve proactively.Fse performed test runs and quality control run without any errors and within acceptable range.No further action required by field service.The aia-900 analyzer is functioning as expected.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The st aia-pack ca 19-9, analyte application manual states the following: limitations of the procedure.St aia-pack ca 19-9 should not be used as a screening test.The results obtained from this assay should be used in conjunction with other data (e.G., symptoms, results of other tests, clinical impressions, therapy, etc.).Using st aia-pack ca 19-9, the highest concentration of ca 19-9 measurable without dilution is 400 u/ml, and the lowest measurable concentration is 1.0 u/ml (assay sensitivity).Although the approximate value of the highest calibrator is approximately 400 u/ml, the exact concentration may be slightly different depending upon the lot.The assay specification, assay range high, should be defined as the upper limit of the assay range, 400 u/ml.Although hemolysis has an insignificant effect on the assay, hemolyzed samples may indicate mistreatment of a specimen prior to assay and results should be interpreted with caution.A ca 19-9 result below the upper reference limit of normal does not indicate the absence of malignancy because patients with histopathologic evidence of pancreatic cancer may have ca 19-9 assay values within the range of normal individuals.Patients who do not secrete blood group lewis antigen do not produce ca 19-9.Conversely, a ca 19-9 assay value exceeding the upper limit reference limit of normal does not necessarily indicate the presence of pancreatic malignancy since a small percentage of healthy individuals and individuals with non-malignant conditions as well as malignancy in other locations than the pancreas may have elevations in ca 19-9 assay results.A ca 19-9 assay value should not be interpreted as absolute evidence for the presence or absence of malignant disease of the pancreas.Certain medications may interfere with assay performance.Specimens from patients taking medicines and/or medical treatment may show erroneous results.All results should be interpreted with respect to the clinical picture of the patient.For a more complete understanding of the limitations of this procedure, please refer to the specimen collection and handling, warnings and precautions, storage and stability, and procedural notes sections in this insert.The aia-900 operator's manual, section 12: flags and error messages states the following: dl flag.A fault occurred in the detector or the substrate feed line.Print and display (rate value) : outputs the measurement values.Print and display (concentration value) : becomes blank.Rs232c output (concentration value) : conforms to the setting that is applicable to the case where no concentration is set in the host.Rs232c output (flag) : a.The most probable cause of the reported event is unknown, could not reproduce reported event.
|