Model Number AIA-360
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/10/2020
Event Type  malfunction  
Manufacturer Narrative
Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.A field service engineer (fse) was at the customer's site to address the reported issue.Fse confirmed the nozzle (needle) was bent.Fse replaced the nozzle and performed adjustments.The customer successfully performed quality control (qc); results passed within the acceptable range.The aia-360 analyzer returned to operation.No further action required by field service.A 13-month complaint history review and service history review through aware date of event for similar complaints was performed for serial number (b)(4).There were no other similar complaints found during the searched period.The most probable cause of the reported event was due to damaged sample nozzle.
 
Event Description
A customer reported a bent sample probe on the aia-360 analyzer.A field service engineer was dispatched to address the reported event, which resulted in delayed reporting of patient results for cardiac troponin i (ctnl2).There was no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-360
Type of Device
AIA-360
Manufacturer (Section D)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA  1058623
Manufacturer (Section G)
TOSOH CORPORATION
shiba-koen first building
3-8-2 shiba
minato-ku, 10586 23
JA   1058623
Manufacturer Contact
bernadette oconnell
shiba-koen first building
3-8-2 shiba
minato-ku, tokyo 10586-23
JA   1058623
MDR Report Key9930636
MDR Text Key261066089
Report Number8031673-2020-00096
Device Sequence Number1
Product Code KHO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other
Type of Report Initial
Report Date 03/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-360
Device Catalogue Number019945
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 03/10/2020
Initial Date FDA Received04/06/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-