Model Number CD3369-40Q |
Device Problem
Delayed Charge Time (2586)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
It was reported that while preparing the implantable cardioverter defibrillator prior to implant procedure, upon interrogation it was noted that the device exhibited delayed charge time.The device was not used.There was no patient involvement.
|
|
Manufacturer Narrative
|
The reported event of long charge time was confirmed via review of the device image but could not be replicated in the laboratory.High voltage charging was performed in the laboratory, and the charge time was within specifications.The device¿s high voltage capacitors were analyzed by its manufacturer and no anomalies were found.
|
|