Catalog Number ASK-05500-KM
Device Problem No Flow (2991)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/12/2020
Event Type  malfunction  
Manufacturer Narrative
Qn#: (b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
 
Event Description
It was reported that the medical agent did not flow at the snaplock adaptor attached to the catheter.Therefore, the catheter and the snaplock adaptor were replaced with a new kit.
 
Event Description
It was reported that the medical agent did not flow at the snaplock adaptor attached to the catheter.Therefore, the catheter and the snaplock adaptor were replaced with a new kit.
 
Manufacturer Narrative
(b)(4).A device history record review was performed on the epidural catheter and snaplock assembly with no relevant findings.The customer reported that the snaplock assembly was blocked.The customer returned one 25ml injection syringe, one flat filter, one snaplock assembly, and catheter.The components were received connected together.The returned components were visually examined with and without magnification.Visual examination of the returned snaplock assembly and flat filter revealed they both components appear typical with no observed defects or anomalies.Visual examination of the returned catheter revealed the catheter appears used.Adhesive material can be seen on the outer extrusion.A functional flow test was performed on the returned sample per amrq-000017 section 7.8; rev.7.The returned epidural catheter was inserted from the proximal end into the returned snaplock assembly until it bottomed out and the snaplock assembly was closed.The components were confirmed to be secured by tugging gently.The snaplock assembly was connected to the lab leak tester (ref (b)(4) and the pressure was increased to 10 psi to establish flow.Water could be seen immediately exiting the distal end of the catheter.The flow rate was measured at 8.4ml/min (c05180), which is within the specification of 1ml/min.No blockages were found.The reported complaint of the snaplock being blocked could not be confirmed based on the sample received.A device history record review was performed on the epidural catheter and snaplock assembly with no evidence to suggest a manufacturing related issue.The returned snaplock assembly passed a functional flow test.There were no functional issues found with the returned sample.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPIDURAL CATHETERIZATION KIT
Type of Device
ANESTHESIA CONDUCTION KIT
MDR Report Key9933819
MDR Text Key187031371
Report Number3006425876-2020-00286
Device Sequence Number1
Product Code CAZ
Combination Product (y/n)N
PMA/PMN Number
K103658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Remedial Action Modification/Adjustment
Type of Report Initial,Followup
Report Date 03/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/15/2021
Device Catalogue NumberASK-05500-KM
Device Lot Number71F19H2205
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2020
Initial Date Manufacturer Received 03/16/2020
Initial Date FDA Received04/08/2020
Supplement Dates Manufacturer Received05/04/2020
Supplement Dates FDA Received05/11/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/26/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-