• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM; PERIPHERAL ATHERECTOMY DEVICE Back to Search Results
Model Number VPR-GW-FLEX14
Device Problem Material Separation (1562)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 03/26/2020
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.The material inspection report for this guide wire lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.(b)(4).
 
Event Description
During a peripheral procedure, the tip of the viperwire guide wire sheared off in the common femoral artery after several wire and catheter exchanges.The target lesion was located in a diseased femoral artery accessed via a contralateral approach.The fragment was post-dilated into the wall of the femoral artery without any obvious or immediate compromise to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM
Type of Device
PERIPHERAL ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul, mn
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old highway 8 nw
saint paul, mn
Manufacturer Contact
sarah hicks
1225 old highway 8 nw
saint paul, mn 
MDR Report Key9948163
MDR Text Key188340318
Report Number3004742232-2020-00117
Device Sequence Number1
Product Code MCW
UDI-Device Identifier10852528005718
UDI-Public(01)10852528005718(17)211130(10)297060
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190634
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2021
Device Model NumberVPR-GW-FLEX14
Device Catalogue Number7-10041-03
Device Lot Number297060
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/27/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/11/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age85 YR
-
-