It was reported that the implantable cardioverter-defibrillator successfully delivered therapy when the patient experienced ventricular fibrillation (vf).Non sustained right ventricular oversensing episodes were observed prior to the shock delivered that showed undersensing on the discrimination channel.No therapies were inhibited.After the shock, there were episodes of a rhythm similar to a bigeminal rhythm, but the signals were not seen on the far field left or right channels.These following signals were likely not true signals from the heart and also no longer happening.The patient was hospitalized with symptoms of cardiac arrest.Programming changes were discussed but could not be performed as the patient passed away on the next day.The device did not deliver therapy for ten minutes and caused the patient to be intubated.The patient never recovered.The device might cause or contribute to patient's death.
|