Model Number 10062818 |
Device Problems
Fluid/Blood Leak (1250); Material Separation (1562)
|
Patient Problem
No Code Available (3191)
|
Event Date 02/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.The requested information in patient demographics was not provided.Per customer, "unable to disclose due to patient confidentiality.".
|
|
Event Description
|
It was reported that during blood transfusion, the flow was running slowly.The nurse had to milk the tubing to help with the flow.However, the tubing came apart at the connection point.There was a delay in treatment as the blood transfusion had to be stopped and a new tubing needed to be set up.
|
|
Manufacturer Narrative
|
Additional information added to: d11 the customer¿s report that the blood tubing separated was confirmed.Further visual inspection of the separated silicone segment end noted o-ring indentations, indicating that the retainer ring had been properly assembled on the set.The silicone pump tubing was found to be within measurement specification.Functional testing could not be performed due to the set's separation.Blood fluid was observed to be noted from the separation.Visual inspection confirmed the set was separated at the pumping segment.The separation was at the engagement between the silicone pump segment tubing and the upper fitment.The ring retainer (p/n 601316-000) was not received with the set.No gaps on the silicone segment separation or any anomalies were noted during visual inspection.Functional testing could not be performed due to the set's separation and obvious leaking that would occur due to the separation.Device history record could not be performed on model 10062818 because the lot # for the suspect set was not provided.Although the root cause of the separation could not be definitively determined, at some point the set may have been pulled or stretched beyond the retention specification between the silicone segment and the set¿s upper fitment during use or set loading.
|
|
Event Description
|
It was reported that during blood transfusion, the flow was running slowly.The nurse had to milk the tubing to help with the flow.However, the tubing came apart at the connection point.There was a delay in treatment as the blood transfusion had to be stopped and a new tubing needed to be set up.
|
|
Search Alerts/Recalls
|