U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problem Component Missing (2306)
Patient Problem No Patient Involvement (2645)
Event Date 01/21/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Product review of the device by flextronics on 17 feb 2020 revealed the control bar was bent, the unit was out of calibration, 2 screws were missing and the needle bearing was corroded.Repair of the device was performed by flextronics on 17 feb 2020 which included replacement of the control bar, shaft bearings, sleeve bearings, needle bearing, and screws.The device, serial number (b)(4), was then tested and functioned properly.It was repaired, inspected and tested.Review of the device history record identified no deviations or anomalies during manufacturing.Device is used for treatment.Review of complaint history identified additional similar complaints for the reported item and part and lot combination.Complaints are monitored through monthly complaint review in order to identify potential adverse trends.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event cannot be confirmed.
 
Event Description
It was reported that the safe position does not work.During an investigation of the device, it was discovered that screws were missing.No adverse events were reported as a result of this malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
christina arnt
56 e. bell drive
warsaw, IN 46582
5745273773
MDR Report Key9962919
MDR Text Key188432209
Report Number0001526350-2020-00376
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 04/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot Number31609500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/22/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/06/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/07/1998
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
-
-