Model Number 60-02-15 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/09/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova (b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.A livanova field service representative was dispatched to the facility to investigate the device and could confirm the reported event.He traced the failure back to a defective processor board.He replaced the part and the issue was solved.Subsequent functional verification testing was completed without further issues and the unit was returned to service.The root cause of the event has been identified as a defective cpu board.
|
|
Event Description
|
Livanova received a report that a centrifugal pump system with tubing clamp displayed an error code associated to a processor error during a procedure and all buttons on the control panel were not working.There was no report of patient injury.
|
|
Manufacturer Narrative
|
H10: through follow-up communication livanova learned that the user hand-cranked the pump until the control panel was quickly replaced with another and the procedure could be completed without further issues.In addition the correct date of the event was (b)(6) 2020.It has been corrected in the dedicated section b3.
|
|
Event Description
|
See initial report.
|
|