This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation noted evidence indicating difficulty may have been encountered fully inserting a is-4 lead into the header of this device resulting in oversensing with some pacing inhibition.Please refer to the description for more information regarding the specific circumstances of this event.
|
It was reported that during this device implant, the physician tried reconnecting the lead into the header several times.Noise oversensing with some pacing inhibition was noted, each time during isometric testing.Set screw issues were also noted.The physician said it did not feel quite right in the header.Thus, a new device was used, with no issues.No adverse patient effects were reported.This product has not been returned.This report will be updated, should additional information be received.
|