Model Number 1458Q/86
Device Problem High Capture Threshold (3266)
Patient Problem Undesired Nerve Stimulation (1980)
Event Date 04/01/2020
Event Type  Injury  
Manufacturer Narrative
This product is registered as a combination product.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
It was reported that a patient experienced phrenic nerve stimulation - pns from their left ventricular lead during implantation.The lead only had one vector that did not cause phrenic stim due to the patient's anatomy.Patient still experienced pns shortly after the implant and was brought in for reprogramming on (b)(6) 2020.It was during this time that the healthcare provider found that the lead also had an increased capture threshold.The programming was able to minimize pns to a manageable level for the patient.The patient experienced an increase in pns in the newly programmed lead shortly after the clinic visit, so the lead was turned off.No new intervention has been performed since.Patient condition was stable after the events.
 
Event Description
New information received notes that the lead was explanted and replaced on (b)(6) 2020.Patient condition was stable after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key9977397
MDR Text Key188218292
Report Number2017865-2020-05138
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot NumberA000088355
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/12/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/23/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-