Model Number N/A
Device Problems Material Integrity Problem (2978); Device Fell (4014)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/11/2020
Event Type  malfunction  
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Once an investigation of the device is completed, a follow-up/final report will be submitted.
 
Event Description
It was reported that the trampeze traction clamp broke and hit a patient in the head, however the patient was unharmed.No adverse events were reported as a result of this malfunction.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).Visual examination of the returned product identified that the product had fractured.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
 
Event Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLAIN BAR 96 IN. (244 CM) TRAPEZE
Type of Device
APPARATUS, TRACTION, NON-POWERED
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
MDR Report Key9981727
MDR Text Key191795911
Report Number0001526350-2020-00391
Device Sequence Number1
Product Code HST
Combination Product (y/n)N
PMA/PMN Number
PREAMENDMENT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00064000300
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/11/2020
Initial Date FDA Received04/21/2020
Supplement Dates Manufacturer Received06/11/2020
Supplement Dates FDA Received06/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age61 YR
Patient Weight110
-
-