The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking catheter was unconfirmed because the problem could not be reproduced.The product returned for evaluation was one 4fr d/l power midline catheter.Usage residues were observed throughout the sample.Microscopic inspection of the sample did not reveal any evidence of damage or prior leakage.An attempt to infuse water through the sample using a 12ml syringe revealed both lumens to be patent to infusion and aspiration with no observed leaks.No leaks were observed during sustained (>15 seconds) hydraulic pressurization of the sample.No deficiencies were discovered during evaluation of the returned sample.Consequently this complaint is unconfirmed at this time.A lot history review (lhr) of redv2088 showed two other similar product complaint(s) from this lot number.The complaints for this lot number (redv2088) have been reported from the same facility.
|