• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL SYRINGE 10ML REG PR SALINE 10ML FILL; SALINE, VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL SYRINGE 10ML REG PR SALINE 10ML FILL; SALINE, VASCULAR ACCESS FLUSH Back to Search Results
Model Number 306546
Device Problem Device Markings/Labelling Problem (2911)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/08/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: no sample was received.No photo was provided; the same day the complaint was reported (4/08/2020) the manufacturing site was informed about it.We confirmed that these products were released for sale on 4/01/2020 and were good for use.The personnel at a distribution center missed to remove the placard from this pallet.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.A device history record review was completed with zero defects found.No quality notifications were written for this batch, nor for the associated assembly batches.Investigation conclusion: this is the 1st complaint for lot # 42292 for this type of defect or symptom.There was no documentation for this type of defect during the entire production run of this batch.Root cause description: the personnel at a distribution center missed to remove the placard from this pallet.Rationale: capa not required at this time.
 
Event Description
It was reported that the syringe 10ml reg pr saline 10ml fill was labeled "defective" and received by customer.Product defect was noted prior to use.The following information was provided by the initial reporter: material no.: 306546 batch no.: 0042292.We received a pallet today that had the below sign on it: qa quarantine.Awaiting test results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYRINGE 10ML REG PR SALINE 10ML FILL
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652341
MDR Report Key9986738
MDR Text Key189197210
Report Number1911916-2020-00401
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public30382903065463
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161552
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date01/31/2023
Device Model Number306546
Device Catalogue Number306546
Device Lot Number0042292
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/08/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/11/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-