Model Number 10101
Device Problem Smoking (1585)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/07/2020
Event Type  malfunction  
Manufacturer Narrative
A field service engineer was dispatched to the customer site.The catalytic converter, oil mist filter, and vacuum pump oil were replaced to resolve the smoke/haze issue.Unit meets specifications and was returned to service.Asp complaint ref #: (b)(4).
 
Event Description
A customer reported an event of smoke or haze emitting from the sterrad® 100s sterilizer.There was no report of any injuries or human reactions.The customer was advised to discontinue use of the unit and leave the room until the smoke cleared.An asp field service engineer was dispatched to assess the unit onsite.This event is being reported as a malfunction report subsequent to a serious injury.
 
Manufacturer Narrative
Asp investigation summary: the investigation included a review of the device history record (dhr), trending analysis of the smoke/haze issue, and system risk analysis (sra).The dhr was reviewed and no issues relating the failure mode were noted.The involved unit met manufacturer specifications at the time of release.Trending analysis of the smoke/haze issue was reviewed within the past six months and no significant trend was observed.The sra shows the risk for exposure to toxic or corrosive material to be "low." the parts were not available for return and further analysis.The assignable cause of the smoke/haze issue is the catalytic converter, oil mist filter, and vacuum pump oil.The field service engineer replaced these parts and confirmed the sterrad® 100s was restored to proper function after service.The issue was resolved at the customer facility.Asp will continue to track and trend this issue.Asp complaint ref #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERRAD 100S STERILIZER
Type of Device
STERRAD® EQUIPMENT
Manufacturer (Section D)
ADVANCED STERILIZATION PRODUCTS
33 technology drive
irvine CA
MDR Report Key9993720
MDR Text Key188987174
Report Number2084725-2020-50063
Device Sequence Number1
Product Code MLR
UDI-Device Identifier10705037014460
UDI-Public10705037014460
Combination Product (y/n)N
PMA/PMN Number
K991999
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 08/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number10101
Device Catalogue Number10101
Device Lot NumberN/A
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 02/07/2020
Initial Date FDA Received04/23/2020
Supplement Dates Manufacturer Received08/10/2020
Supplement Dates FDA Received08/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-