ADVANCED BIONICS, LLC HIRES 90K IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1400-01
Device Problems Inaudible or Unclear Audible Prompt/Feedback (2283); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Undesired Nerve Stimulation (1980); No Known Impact Or Consequence To Patient (2692)
Event Date 08/25/2020
Event Type  Injury  
Event Description
The recipient reportedly experienced poor performance and facial nerve stimulation.Programming adjustments were made, and the facial nerve stimulation improved, however, the recipient continued to experience sound quality issues.Revision surgery will be scheduled.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient's issues have reportedly resolved and device activation was successful.The external visual inspection revealed the array was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented some electrical test from being performed.The device passed the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This version of the hires device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
The recipient's device was reportedly explanted.The recipient was reimplanted with another cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key9995145
MDR Text Key190007437
Report Number3006556115-2019-00836
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup
Report Date 10/22/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/01/2012
Device Model NumberCI-1400-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2020
Initial Date Manufacturer Received Not provided
Initial Date FDA Received04/24/2020
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received10/07/2020
12/12/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age14 YR
-
-