Catalog Number 157001135 |
Device Problems
Degraded (1153); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Fatigue (1849); Foreign Body Reaction (1868); Inflammation (1932); Nerve Damage (1979); Pain (1994); Loss of Range of Motion (2032); Anxiety (2328); Distress (2329); Discomfort (2330); Injury (2348); Limited Mobility Of The Implanted Joint (2671); Test Result (2695); No Code Available (3191); Metal Related Pathology (4530)
|
Event Date 01/01/2018 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|
Event Description
|
Patient initiated complaint.It was reported that the patient experienced constant pain and recent test show that he have elevated metal ion levels in the blood.Patient is scheduled for a revision in (b)(6) this year and he asking for a compensation.Doi: 2007.Dor: not revised; (unknown hip).
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report. added: a2 (age), d1, d2, d4 (procode,lot,udi) h4 and d11.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.H10 additional narrative: added: b5, b7, h6 (patient).Corrected: b1, b2, h6 (device).
|
|
Event Description
|
After review of medical records, patient was revised to address mechanical loosening and pain due to recalled right total hip arthroplasty.Revision notes reported that the blood testing demonstrate elevated serum cobalt chromium.Upon going to the capsule, it was noted that there was a significant amount of metal stained tissue.The muscle and primary tendons were not damaged by the metallosis.There was some corrosion at the trunnion.Radiographically, the acetabular component appeared to be loose with radiolucent lines in all zones.Doi: (b)(6) 2007.Dor: (b)(6) 2018.Right hip.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: d6b and h6 (impact code).Removed date of explant and impact code device revision or replacement.
|
|
Event Description
|
Plaintiff fact sheet received.In addition to what was previously alleged.Pfs alleged fatigue, sciatica, plantar fasciitis, difficulty sleeping, metallosis, mental health problem, fear, anxiety and limited adl.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Search Alerts/Recalls
|
|