• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOPERSURGICAL, INC. DONUT PESS 3-3/4 IN #7

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOPERSURGICAL, INC. DONUT PESS 3-3/4 IN #7 Back to Search Results
Model Number MXPDO07
Device Problems Material Discolored (1170); Difficult to Insert (1316); Use of Device Problem (1670); Material Integrity Problem (2978); Physical Resistance/Sticking (4012)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/23/2020
Event Type  malfunction  
Manufacturer Narrative
Coopersurgical inc.Is currently investigating the reported complaint condition.Ref: (b)(4).
 
Event Description
Report stated - "issue with each of them pertaining to texture (to hard), color, stiffness, ease of use.The customer has had ongoing issue with the listed pessaries not being able to insert." ref (b)(4).Donut pess 3-3 4 in no 7 mxpdo07 (b)(4).
 
Manufacturer Narrative
Investigation: x-initiated manufacturer's investigation.X-review dhr.X-inspect stock product.Analysis and findings: ref: (b)(4).Distribution history: the complaint product was manufactured at csi in october 2019 under work order (b)(4).Manufacturing record review.(b(4) was reviewed and no non-conformities, related to the complaint condition, were noted.Historical complaint review: a review of the 2-year complaint history shows one other complaint for this issue.Root cause quality engineering, manufacturing, marketing, and product surveillance conducted an investigation into the recent complaints regarding the hardness and rigidity of the pessary product line.There have been no specification changes in the last 2 years.Returned product from historical complaint investigations and sample product from finished goods inventory were evaluated and the reported condition could not be duplicated.The products were found to meet all approved release specifications.It should be noted that this product was originally manufactured for coopersurgical by a supplier (eis) until mid-2017.Manufacturing was moved to coopersurgical and validated in november 2017 (val-16-0219) using a new silicone supplier.It is suspected the customer noticed the current production pessary was slightly harder than the previously received supplier pessary.It should also be noted that the current silicone supplier provides certifications on every shipment with durometer hardness testing per astm d2240 which verifies durometer specifications are met.Based upon an analysis of historical returned/complaint product, stock product, dhr review and certificates of conformance, the reported complaint condition could not be confirmed.Corrective action: capa 735 has been opened to determine if customer observations with regard to pessary durometer warrant further product enhancements.
 
Event Description
Report stated - "issue with each of them pertaining to texture (to hard), color, stiffness, ease of use.The customer has had ongoing issue with the listed pessaries not being able to insert.".Ref (b)(4).1216677-2020-00109 donut pess 3-3 4 in no 7 mxpdo07 (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DONUT PESS 3-3/4 IN #7
Type of Device
DONUT PESS 3-3/4 IN #7
Manufacturer (Section D)
COOPERSURGICAL, INC.
95 corporate drive
trumbull, ct
MDR Report Key9998020
MDR Text Key191533453
Report Number1216677-2020-00109
Device Sequence Number1
Product Code HHW
Combination Product (y/n)N
PMA/PMN Number
K904774
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 07/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMXPDO07
Device Catalogue NumberMXPDO07
Device Lot Number251267
Was Device Available for Evaluation? Yes
Date Manufacturer Received03/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-