Lot Number 36887121
Device Problem High Test Results (2457)
Patient Problem Test Result (2695)
Event Date 10/29/2018
Event Type  Injury  
Event Description
Inr testing date (b)(6) 2018 per doctor office instructions with home inr monitoring (alere).Home test strip result 4.5 using test strip lot #29415123.Went to hosp, lab result 3.5, therefore instructions were to keep same dosage as had been on.
 
Event Description
Add'l info rec'd from reporter on 02/13/2019 for report number mw5081132: reporter stated that she took her inr test and the result was 4.7 and she took her test again in a little while and it was 4.5.She was told to take coumadin and go to the lab in the morning and the test result was 2.6.She said her doctor told her to inject lovenox if her inr is below 3.The treatment has been withheld based on the result she rec'd.Her doctor told her to go back to the lab on (b)(6) 2019 for follow up.Other#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COAGUCHEK XS PT TEST STRIP
Type of Device
TEST, TIME, PROTHROMBIN
Manufacturer (Section D)
ROCHE DIAGNOSTICS GMBH
MDR Report Key8046468
MDR Text Key126664584
Report NumberMW5081132
Device Sequence Number2
Product Code ---
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 02/13/2019
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator No Information
Device Expiration Date05/31/2020
Device Lot Number36887121
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/06/2018
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/13/2019
Was Device Evaluated by Manufacturer? No Information
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age62 YR
Patient Weight65
-
-