• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN SORIN 3T HEATER COOLER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN SORIN 3T HEATER COOLER Back to Search Results
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Information (3190)
Event Date 11/11/2015
Event Type  malfunction  
Event Description
(b)(6) uses sorin 3t heater coolers during cardiac bypass procedures at (b)(6).On (b)(6) 2015 received field safety notice for the heater-cooler system 3t disinfection process, usa from the sorin group, medical technology.Sorin theater cooler machines are cleaned per manufacturer's recommendations.Biotest labs (the lab recommended for sending surveillance cultures to by sorin) notified us of some growth on samples sent from (b)(6) 2015.At our request, biotest forwarded these specimens to (b)(4) who performed genetic testing to determine if any of the organisms isolated are mycobacteria.Four of the 5 sorin t machines had specimens identified with mycobacterium avium.Sorin 3t heater cooler model# 16-02-85; s/n (b)(4), manuf.Date 08/09; s/n (b)(4), manf.Date 05/11; s/n (b)(4), manuf.Date 12/11; s/n (b)(4) manuf.Date 08/12.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SORIN 3T HEATER COOLER
Type of Device
HEATER COOLER
Manufacturer (Section D)
SORIN
MDR Report Key5454978
MDR Text Key38863167
Report NumberMW5060415
Device Sequence Number3
Product Code DWC
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/18/2016
4 Devices were Involved in the Event: 1   2   3   4  
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Other Device ID Number12/01/2011
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/18/2016
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
-
-