Patient underwent pacemaker placement on (b)(6) 2017.On (b)(6) 2017, patient coded and expired.Autopsy results are pending and vendor for the pacemaker is currently interrogating the machine.At this time, there is no conclusive reason attributing the patient's death to the pacemaker.However, as a cautionary measure, this report is being submitted.Pacemaker: medtronic advisa mri dr ipg, model no.A2dr01, right atrial lead: model 407605, serial # (b)(4), right ventricular lead: model # 407652, serial # (b)(4).
|