• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER IRELAND OSTEONICS TRIDENT PSL ACETABULAR SHELL ; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL / POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER IRELAND OSTEONICS TRIDENT PSL ACETABULAR SHELL ; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL / POLYMER CEMENTED Back to Search Results
Device Problem Corroded (1131)
Patient Problems Inflammation (1932); Pain (1994); Toxicity (2333)
Event Date 10/28/2019
Event Type  Injury  
Event Description
On (b)(6) 2008, bf received a right tha consisting of a stryker lfit, anatomic v40 cobalt-chrome femoral head size 36mm +5mm offset fitted on a accolade tmzf plus hip stem, paired with a trident psl acetabular shell fixed by two bone screws and a neutral trident x3 polyethylene liner.He is a skier and hiker and did not note much problems at the right artificial hip until about spring of 2019, when he developed new activity related right hip and groin pain especially with right hip flexion against resistance and right hip abduction against resistance.In 2015, he was seen by a rheumatologist for right hand weakness due to an undetermined condition that was deemed "rheumatoid-like".His cobalt levels were first checked on (b)(6) 2018, and his urine cobalt level was elevated at 14.6ppb and his whole blood cobalt level was elevated of 4.4 ppb.He began taking n-acetyl cysteine 600-1,800 mg/day for heavy metal chelation on (b)(6) 2018.His levels were rechecked on (b)(6) 2019, and they were 5.3 ppb in blood and 29.3ppb in urine.By this time, he was also noticing some hip symptoms.Metal suppression mri of the right hip showed two discrete foreign body granulomas in the anterior and posterior capsular area of the right hip.Hip abductor tendons look intact.His fdg pet brain scan is consistent with brain hypometabolism.By (b)(6) 2019, his blood cobalt was 4.6 ppb and his urine cobalt was 42ppb.Given the increasing symptoms at the hip, his fdg pet brain scan findings, and his rising cobalt levels, he elected to have the right hip revised on (b)(6) 2019.The trunnion and head bone showed gross corrosion.Posterior, anterior capsule and hip abductor tendons were inflamed.Fluid from right hip was aspirated and diluted 1:2 with local anesthetic, this was sent for cobalt analysis and was found to have a cobalt level of 470ppb.Fda safety report id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT PSL ACETABULAR SHELL
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL / POLYMER CEMENTED
Manufacturer (Section D)
STRYKER IRELAND OSTEONICS
MDR Report Key9391811
MDR Text Key168588383
Report NumberMW5091353
Device Sequence Number3
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/26/2019
6 Devices were Involved in the Event: 1   2   3   4   5   6  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator No Information
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/27/2019
Was Device Evaluated by Manufacturer? No Information
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age64 YR
Patient Weight98
-
-