Model Number 2274-63-000
Device Problems Break (1069); Material Twisted/Bent (2981); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2019
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: the examination of the returned instrument confirmed the complaint.Evaluation of the returned screwdriver found the tip is stripped and the screwdriver was fractured in the distal forked component at the pin junction of the forked hex screwdriver tip to the swivel block.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the depth gauge broke in 2 pcs right at the bend.No surgical delay.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint #: (b)(4).Investigation summary: the examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: null.Device history batch: null.Device history review: null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET TPRD HEX SCDR U-JOINT
Type of Device
HIP INSTRUMENTS : SCREWDRIVERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10002355
MDR Text Key189089970
Report Number1818910-2020-11390
Device Sequence Number1
Product Code HXX
UDI-Device Identifier10603295109365
UDI-Public10603295109365
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2274-63-000
Device Catalogue Number227463000
Device Lot NumberSO2032782
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/06/2020
Date Manufacturer Received04/24/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-