• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR MITRACLIP DELIVERY SYSTEM; VALVE REPAIR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR MITRACLIP DELIVERY SYSTEM; VALVE REPAIR Back to Search Results
Model Number CDS0601-NTR
Device Problems Unstable (1667); Mechanical Jam (2983)
Patient Problem No Patient Involvement (2645)
Event Date 04/06/2020
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history did not identify any similar incidents reported from this lot.The investigation determined the reported out of socket arm positioner result in mechanical jam appears to be related to a potential product issue.The issue is being addressed per internal operating procedures.Abbott will continue to trend the performance of these devices.
 
Event Description
This is filed to report the unstable arm positioner.It was reported that during preparation, the back end of the arm positioner looked to be out of socket and was unable to rotate.Therefore, the clip delivery system (cds) was not used and was replaced.There was no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP DELIVERY SYSTEM
Type of Device
VALVE REPAIR
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key10005441
MDR Text Key189224941
Report Number2024168-2020-03946
Device Sequence Number1
Product Code NKM
UDI-Device Identifier08717648226342
UDI-Public08717648226342
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/02/2020
Device Model NumberCDS0601-NTR
Device Catalogue NumberCDS0601-NTR
Device Lot Number91203U148
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/06/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/03/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-