Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed that the electrode was severed prior to receipt.In addition, severe surgical tool damage was observed on the top and bottom cover and seam weld.These anomalies are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented some electrical tests from being performed.The device passed some of the electrical tests performed.The scanning electron microscopy analysis revealed a crack on the seam weld which is believed to have been sustained during the revision surgery.The internal visual inspection noted silver migration across some electrical components.The reported complain of overly loud sound could not be verified during this analysis, which was limited in some respects due to the damage the device sustained during the revision surgery.This device had a gross leak failure at the case-band braze, which is believed to have been induce during revision surgery.Due to the state of the device most tests were compromised and pinpointing the root cause was not possible.A corrective action was implemented.Feed thru assemblies from his vendor are no longer used.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|