• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUICKFLEX MICRO LV LEAD; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUICKFLEX MICRO LV LEAD; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1258T/86
Device Problems Signal Artifact/Noise (1036); Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/20/2020
Event Type  Injury  
Manufacturer Narrative
This product is registered as a combination product.The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Related manufacturer reference number: 2938836-2020-02789.It was reported that the right and left ventricular lead exhibited noise.During the procedure, insulation abrasion with extent fillars were visibly confirmed on the leads to can in the pocket.Both leads were explanted and replaced.No signs or symptoms of infection were noted.The patient was stable before, during and after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKFLEX MICRO LV LEAD
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key10016335
MDR Text Key189575732
Report Number2938836-2020-02790
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734502610
UDI-Public05414734502610
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2016
Device Model Number1258T/86
Device Catalogue Number1258T-86
Device Lot Number4262353
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/20/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/24/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
TENDRIL STS
Patient Outcome(s) Required Intervention;
Patient Weight75
-
-