Model Number 1458Q/86
Device Problem Failure to Capture (1081)
Patient Problem Cardiopulmonary Arrest (1765)
Event Date 04/18/2020
Event Type  Injury  
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.This product is registered as a combination product.
 
Event Description
Related manufacturer reference number: 2017865-2020-05806, related manufacturer reference number: 2017865-2020-05813.It was reported that the patient presented as unresponsive during an in-patient hospital stay; subsequent resuscitation was required.Upon interrogation of the implantable cardioverter defibrillator recorded episodes of non-sustained right ventricular over-sensing, were revealed.There appeared to be intermittent loss of capture exhibited by both the right and left ventricular leads.Programming interventions were made.The patient was stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key10033295
MDR Text Key190158488
Report Number2017865-2020-05810
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 05/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2017
Device Model Number1458Q/86
Device Catalogue Number1458Q-86
Device Lot Number4818127
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received04/18/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/03/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
QUADRA ASSURA CRT-D QUAD RF HV; RIATA ST ACTIVE FIXATION
Patient Outcome(s) Life Threatening;
Patient Age76 YR
-
-