Catalog Number CS-15552-VSP
Device Problems Crack (1135); Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/20/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The customer reports the hub of the catheter has a big crack on it and it causes leakage.The blue part is cracked right through.The hub was replaced.Catheter was placed late (b)(6) 2020.
 
Manufacturer Narrative
Qn#(b)(4).The sample was not returned; however, the customer did provide one photo of an extension line hub.The photo confirmed that the blue luer hub contained a crack.A complete visual inspection could not be performed as no sample was returned for analysis.The customer did not provide a lot number; therefore, a device history record review was performed based upon a lot number from the sales history data of the customer.No relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "repeated over tightening of bloodlines, syringes and caps will reduce connector life and could lead to potential connector failure." the complaint of a cracked luer hub was confirmed by the customer returned photo.Complete complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed based on sales history, and no relevant findings were identified.Without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.If the sample becomes available at a later date a follow up report will be submitted with investigation results.
 
Event Description
The customer reports the hub of the catheter has a big crack on it and it causes leakage.The blue part is cracked right through.The hub was replaced.Catheter was placed late (b)(6) 2020.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW HEMODIALYSIS KIT: 2-L 15 FR X 50 CM
Type of Device
CATHETER, HEMODIALYSIS, IMPLA
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key10033448
MDR Text Key190165386
Report Number9680794-2020-00219
Device Sequence Number1
Product Code MSD
Combination Product (y/n)N
PMA/PMN Number
K111117
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/07/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCS-15552-VSP
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Date Manufacturer Received06/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age56 YR
-
-