EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS; HEART-VALVE, REPLACEMENT
|
Back to Search Results |
|
Model Number 3300TFX |
Device Problems
Inadequacy of Device Shape and/or Size (1583); Appropriate Term/Code Not Available (3191)
|
Patient Problems
Aortic Regurgitation (1716); Aortic Valve Stenosis (1717); No Known Impact Or Consequence To Patient (2692)
|
Event Date 04/15/2020 |
Event Type
Injury
|
Manufacturer Narrative
|
Patient prosthesis mismatch (ppm) is present when the effective orifice area of the inserted prosthetic valve is too small in relation to body size.Its main hemodynamic consequence is to generate higher than expected gradients through a normally functioning prosthetic valve.Ppm has been shown to be associated with worse hemodynamic function, less regression of left ventricular hypertrophy, more cardiac events, and lower survival.The device was not returned for evaluation, as it remains implanted.The root cause of this event cannot be conclusively determined.However, it is likely that patient related and/or procedural factors contributed to the event.The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
|
Event Description
|
It was reported that a 21mm pericardial aortic valve was disabled via a valve-in-valve procedure on pod #49 due to undersizing, stenosis, regurgitation, and pvl.A 23mm transcatheter valve was implanted.Patient was discharged to inpatient rehab in fair condition on pod #7.As reported, the physician did not allege malfunctions of the surgical valve.
|
|
Manufacturer Narrative
|
Additional manufacturer narrative.Updated h6 per new information received.
|
|
Manufacturer Narrative
|
Corrected data: f10, h6.Reference capa-20-00141.
|
|
Search Alerts/Recalls
|
|
|