Model Number N/A |
Device Problems
Flaked (1246); Peeled/Delaminated (1454)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/25/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source: (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The product remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that a metal fragment flaked off of the taperloc stem during insertion of implant.No additional information is available.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
(b)(4).Reported event was confirmed by review of photograph of device fracture.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.No product was returned.A picture of the fractured piece was provided.It appears to be possible part of porous coating.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|