ST PAUL LEVEL 1 HOTLINE LOW FLOW SYSTEM; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-90
Device Problem Noise, Audible (3273)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
Evaluation results: one level 1 hotline low flow system (h-l90) was returned for investigation in used condition.Visual inspection revealed that the unit had an old style front cover, enclosure and water tank cover.There was a crack on the bottom of the enclosure, as well as on the front cover and water tank cover.The unit was also equipped with and older line cord and pole mount.A power on test was performed and the customer reported product problem (loud noise) was confirmed.The product problem was attributed to the pump which was over 10 days old.The pump was replaced as result.The aforementioned damaged/worn components were also replaced.The device subsequently passed all the functional tests.
 
Event Description
It was reported that the fluid warmer produced loud noises.No patient injury or complications were reported in relation to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE LOW FLOW SYSTEM
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key10063220
MDR Text Key191181731
Report Number3012307300-2020-04447
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier50695085820909
UDI-Public50695085820909
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 05/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberHL-90
Device Catalogue NumberHL-90
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/11/2020
Initial Date FDA Received05/15/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/27/1998
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-