Model Number 746F8
Device Problems Inflation Problem (1310); Material Rupture (1546)
Patient Problems No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Event Date 04/10/2020
Event Type  malfunction  
Manufacturer Narrative
One catheter with attached monoject 1.5 cc limited volume syringe was returned for evaluation.Balloon did not inflate due to leakage from a tear, approximately 0.5 mm long, along the distal balloon bonding site.The edges of the balloon latex at the tear appeared to be different shapes and were not able to match up.All through lumens were patent without any leakage or occlusion.No other visible damage or abnormality was observed from the catheter body or returned syringe.Balloon inflation test was performed using returned syringe with 1.5 cc air by holding the balloon under water.Visual examination was performed under microscope at 20x magnification.A device history record review was completed and documented that device met all specifications upon distribution.Customer report of balloon inflation issue was confirmed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.It is common clinical practice to check balloon integrity by inflating it to the recommended volume in order to detect any asymmetry or leakage condition before use of the catheter.It is unknown if user or procedural factors may have contributed to the event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the balloon of the swan ganz catheter did not inflate before use.There were no patient complications reported.Patient demographic information requested but unavailable.
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO CCO/SVO2/VIP THERMODILUTION CATHETER
Type of Device
SWAN-GANZ CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
MDR Report Key10063513
MDR Text Key194634826
Report Number2015691-2020-11879
Device Sequence Number1
Product Code DYG
Combination Product (y/n)N
PMA/PMN Number
K934742
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 04/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/07/2021
Device Model Number746F8
Device Catalogue Number746F8
Device Lot Number62622173
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2020
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-