BARD ACCESS SYSTEMS HGSTAR, 14.5 FR, ST, 19CM, STAND; DIALYSIS CATHETER Back to Search Results
Model Number 5833690
Device Problems Separation Failure (2547); Difficult or Delayed Separation (4044)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/10/2020
Event Type  malfunction  
Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation, however photos have been provided.The investigation of the reported event is currently underway.(expiration date: 12/2020).
 
Event Description
It was reported that during a catheter placement procedure through right subclavian vein, the introducer sheath allegedly could not be peeled properly.The procedure was completed using another device.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: the device history record review was performed for the reported lot number and this lot meets all release criteria.The manufacturing review did not indicate any possible manufacturing issue that could be related to the reported event.Investigation summary: the sample was not returned for evaluation, however, one electronic photo was provided for review.The investigation is confirmed for failure to separate the sheath, as the sheath in the photo was not completely peeled and still intact near the distal end of the sheath.A definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that during a catheter placement procedure through right subclavian vein, the introducer sheath allegedly could not be peeled properly.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HGSTAR, 14.5 FR, ST, 19CM, STAND
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key10074879
MDR Text Key191565054
Report Number3006260740-2020-01824
Device Sequence Number1
Product Code MSD
UDI-Device Identifier00801741013386
UDI-Public(01)00801741013386
Combination Product (y/n)N
PMA/PMN Number
K051748
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5833690
Device Catalogue Number5833690
Device Lot NumberREDN4087
Was Device Available for Evaluation? No
Date Manufacturer Received10/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age65 YR
Patient Weight55
-
-