Model Number DX115
Device Problem Component Incompatible (1108)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Investigation results: there are discolorations and residues remained at the transition between metal part and silicone.The components were examined visually and microscopically.Under microscopic enlargement it can be seen that the metal part is covered with a thin transparent silicon layer.This layer shows several damages.Several residues/contaminations have been deposited in this area.Inspection of unused stored goods: the stored goods show also a thin silicon layer which covers the metal part.This silicon layer shows also partly damages.At this stage of investigation we assume a potential manufacturing error therefore the complaint was forwarded to the manufacturer, as this product is manufactured by third party.Up to now the investigation/statement from the manufacturer is still pending.The product does not require batch management; a review of the device quality and manufacturing history records is not possible.Due to the circumstance that a statement/investigation from the manufacturer side still await it is not possible to determine a root cause for the mentioned failure.The investigation is ongoing.An internal inspection was initiated ((b)(4)).
 
Event Description
It was reported that there was an issue with a oral silicone wedge.According to the customer it was reported that the product cannot be cleaned properly.During a routine check the customer found residues between the metal insert and the silicone (protein test was positive).There was no patient harm.Additional information was not provided nor available.The malfunction is filed under ((b)(4)).Associated medwatch-reports: 9610612-2020-00175 ((b)(4) - dx113).
 
Manufacturer Narrative
Additional case created.See b5.
 
Event Description
Additional product with the notification number (b)(4) was created.Awareness date is 04.28.2020.Associated medwatch report: 9610612-2020-00175 ((b)(4) dx113); 9610612-2020-00176 ((b)(4) dx115).
 
Manufacturer Narrative
Investigation results: there are discolorations and residues remained at the transition between metal part and silicone.Visual investigation: under microscopic enlargement it can be seen that the metal part is covered with a thin transparent silicon layer.This layer shows several damages (tears).Several residues/contaminations have been deposited in this area.At this stage of investigation we assume a potential manufacturing error therefore the complaint was forwarded to the manufacturer, as this product is manufactured by third party.Up to now the statement from the manufacturer is still pending.Batch history review: the product does not require batch management; a review of the device quality and manufacturing history records is not possible.Conclusion and measures: at that time it is not possible to determine a clear conclusion and root cause for the mentioned failure.Due to the circumstance that a statement/investigation from the manufacturer side still await it is not possible to determine a root cause for the mentioned failure.The investigation is ongoing.As a preventive measure a internal inspection was opened under psc 2020-020 and since no statement of the manufacturer is available, the articles are no longer distributed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ERGOPLANT ORAL SILICONE WEDGE #L
Type of Device
DENTAL
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key10074926
MDR Text Key194009463
Report Number9610612-2020-00176
Device Sequence Number1
Product Code DZN
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Remedial Action Inspection
Type of Report Initial,Followup,Followup
Report Date 05/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDX115
Device Catalogue NumberDX115
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2020
Date Manufacturer Received05/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-