EDWARDS LIFESCIENCES CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS; HEART-VALVE, REPLACEMENT Back to Search Results
Model Number 3300TFX
Device Problems Inadequacy of Device Shape and/or Size (1583); Patient-Device Incompatibility (2682)
Patient Problems Aortic Valve Stenosis (1717); Stenosis (2263)
Event Date 04/27/2020
Event Type  Injury  
Manufacturer Narrative
Patient prosthesis mismatch (ppm) is present when the effective orifice area of the inserted prosthetic valve is too small in relation to body size.Its main hemodynamic consequence is to generate higher than expected gradients through a normally functioning prosthetic valve.Ppm has been shown to be associated with worse hemodynamic function, less regression of left ventricular hypertrophy, more cardiac events, and lower survival.The patient may require explant of the device or intervention may be indicated.The device was not returned for evaluation, as it remains implanted.The root cause of this event cannot be conclusively determined.However, it is likely that patient related factors and the progression of the underlying valvular disease pathology contributed to the event.  the device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
It was reported that a 21mm pericardial aortic valve was disabled via a valve-in-valve procedure after an implant duration of 6 years due to severe stenosis and undersizing.A 23mm transcatheter valve was implanted.Patient tolerated well and is recovering post procedure without complication.Patient was discharged.As reported, the physician did not directly allege malfunction of the surgical valve.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS
Type of Device
HEART-VALVE, REPLACEMENT
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
MDR Report Key10079683
MDR Text Key191731675
Report Number2015691-2020-11919
Device Sequence Number1
Product Code DYE
Combination Product (y/n)N
PMA/PMN Number
P860057/S042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 04/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/21/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/23/2018
Device Model Number3300TFX
Was Device Available for Evaluation? No
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age81 YR
-
-