• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SPIRIT HYDROCOLLOID ADHESIVE SHEATH; MALE EXTERNAL CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SPIRIT HYDROCOLLOID ADHESIVE SHEATH; MALE EXTERNAL CATHETER Back to Search Results
Model Number 35303
Device Problems Nonstandard Device (1420); Difficult to Remove (1528); Material Deformation (2976)
Patient Problems Pain (1994); No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the cunningham clamp should be more circular as it caused pain to the patient and the external catheters were hard to remove.
 
Event Description
It was reported that the cunningham clamp should be more circular as it caused pain to the patient and the external catheters were hard to remove.
 
Manufacturer Narrative
The reported event was inconclusive.It was unknown whether the device had met specifications due to no sample was returned for evaluation.The product was used for treatment purposes.A potential failure mode could be ¿too high viscosity¿ with a potential root cause of ¿viscometer failure or mechanical failure¿.This failure falls in a quad 2 risk region and based on this information the risk is moderate.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warm water.Dry thoroughly.2) trim pubic hair if necessary.3) unroll self-adhering catheter over penis.4) gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.5) connect to drainage device.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPIRIT HYDROCOLLOID ADHESIVE SHEATH
Type of Device
MALE EXTERNAL CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key10093748
MDR Text Key193044249
Report Number1018233-2020-03464
Device Sequence Number1
Product Code NNX
UDI-Device Identifier00801741070945
UDI-Public(01)00801741070945
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor,other
Type of Report Initial,Followup
Report Date 07/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number35303
Device Catalogue Number35303
Was Device Available for Evaluation? No
Date Manufacturer Received06/15/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-