• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number 42055150-120
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 04/30/2020
Event Type  malfunction  
Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event Description
During preparation inspection of a 5.5x150 supera self expanding stent it was noted that the catheter tip was broken.There was no patient involvement.Another supera was used to complete the procedure.No additional information was provided.
 
Manufacturer Narrative
Visual inspection was performed on the returned device and the reported broken tip was unable to be confirmed.However, noted inner member damage is likely due to inadvertent mishandling during preparation.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other similar incidents reported for this lot.The investigation was unable to determine the cause for the reported broken tip.It is possible that the tip caught on the dispenser coil/packaging or was inadvertently pulled during removal from the packaging causing the noted damage.As such, the noted inner member was stretched and kinked.However, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Type of Device
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10094381
MDR Text Key193603631
Report Number2024168-2020-04555
Device Sequence Number1
Product Code NIP
Combination Product (y/n)N
PMA/PMN Number
P120020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2021
Device Catalogue Number42055150-120
Device Lot Number9120461
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2020
Date Manufacturer Received06/05/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-