Catalog Number 66801196
Device Problem Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 05/07/2020
Event Type  malfunction  
Event Description
It was reported that when the carrier was removed, the silicone adhesive did not remain on the film and removed with the carrier from the film, so it was impossible to use.Backup opsite flexifix gentle was used for treatment.No delay was reported.The sample will be returned for investigation.
 
Manufacturer Narrative
H3, h6: the device intended for use in treatment has been returned and evaluated establishing a relationship between the reported event.The evaluation confirmed silicone remained on the carrier reducing adherence.Root cause determined as a raw material issue.The manufacturing records show no evidence that the product did not meet specification at the time of manufacture.The complaint history file contains further instances of the reported event.This investigation is now complete with no further action deemed necessary at this stage.Smith + nephew are taking further actions relating to the failure reported and continue to monitor for adverse trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPSITE FLEXIFIX GENTLE 5CMX5M
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key10115871
MDR Text Key193719411
Report Number8043484-2020-00620
Device Sequence Number1
Product Code KGX
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,user facilit
Type of Report Initial,Followup
Report Date 03/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/29/2021
Device Catalogue Number66801196
Device Lot Number1353001931
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2020
Distributor Facility Aware Date05/07/2020
Date Manufacturer Received03/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-